Application Area:
• HPLC,UPLC,LC-MS
• ICP-MS,ICP-AES,AAS,GC-MS
• MALDI-TOF-MS,IC,TOC analysis
• Electrochemical, spectrophotometric determination
• Preparation of microbial media and reagents
• Cell culture, PCR, IVF
• Protein purification, electrophoresis, biochemistry
• Proteomics, genomics, immunoassay
• Feed water of laboratory instruments, such as: autoclave, bottle washing machine,
environmental test chamber, water bath, etc.
Main Features:
1.Powerful system overview diagram
•Dynamically display process flow, cartridges configuration,
water quality parameters, RO rejection rate, cartridges life and
tank storage status, etc.
•Button of RO flushing, UF flushing[1] and system disinfection
function can start the forced cleaning of RO membrane, UF
membrane and the system disinfection with chemical dosing,
and remaining time of 3 procedures above is real-time display.
The flushing function significantly prolongs the service life of
the RO membrane and UF membrane, and the disinfection
function effectively sterilizes the pure water pipeline system.
•Ultrapure water circulation system with adjustable interval
running times, to keep the system in low levels of bacterial
contamination and reduce energy consumption.
2.Flexible and diverse water dispensing mode
•With DC24V as the main power supply, fully use weak current
components, to meet the safety standards.
•3 level permission management, administrator and ordinary
users have strict permission distinction.
•Integrated double water leakage protection and alarm device, to
monitor water leakage inside the machine and on the desktop.
•With alarm protection of No feed water, low inlet water
pressure, system high pressure and full tank.
•With alarm function of standard-exceeding of Feed/RO/DI/UP
water quality, and end of cartridge life.
•All alarm information can be stored in the host and cloud
platform, to meet data security requirements.
3.Rigorous double RO system
• Double RO system can remove up to 99% soluble
inorganic ions, 99% soluble organics, microorganisms
and particles.
• Compared with single RO system, the double RO
water quality can be stable < 5μs/cm (feed water
conductivity < 1500μs/cm), and the life of the
ultrapure unit is longer.
• Equipped with DuPont RO membrane, to achieve
combination of long life, stability and high ion
rejection rate.
• Auto-flushing function of RO module with adjustable
flushing interval and duration, to effectively prevent
scale and prolong the life of the membrane.
• The automatic discharge function of unqualified RO
water can ensure that the RO water quality is suitable
to enter the back-end module.
• Integral package of discarded RO module, easy to
install and maintain.
4.Professional mini EDI module
Wi thou t so f tene r and chemical
regeneration, pure water in grade
Ⅱ , with resistivity > 10MΩ.cm@25 ℃
(generally above 15MΩ.cm), and
TOC<30ppb[2], is available. It can
prolong the life of back-end module
and reduce the running cost.